The Guth Model 12V500 Alcohol Breath Simulator is a State-of-the-Art, electronically temperature controlled, water-alcohol instrument for the purpose of providing a precise calibration standard for testing or calibration of alcohol breath analyzers.
Guth型号12V500酒精呼气模拟器是一个先进的电子温度控制,水酒精仪器的目的是为酒精呼气分析仪的测试或校准提供一个精确的校准标准。
The Guth Model 12V500 Alcohol Breath Simulator is a State-of-the-Art, electronically temperature controlled, water-alcohol instrument for the purpose of providing a precise calibration The Guth Model 12V500 Alcohol Breath Simulator is a State-of-the-Art, electronically temperature controlled, water-alcohol instrument for the purpose of providing a precise calibration standard for testing or calibration of alcohol breath analyzers. The Model 12V500 is the first simulator of its kind to incorporate dual temperature probes to independently control and monitor the temperature of the solution. The temperature of the water-alcohol solution placed in the simulator is maintained at 34°C ± .05°C or better. The inlet and outlet port pathways have been designed to be an integral part of the one piece heated aluminum top housing thus allowing uniform heating and reduction of condensation. The new LCD graphics display has a temperature resolution of .01° C and in the event of an error allows feedback to the user in the form of messages instead of codes. As with all Guth products the 12V500 has been engineered using the highest quality components to give reliable, accurate and trouble free operation.
US and Foreign Patents
PERFORMANCE
Temperature Stability: better than ± .05°C
Warm-up Time, Solution and Head Space: less than 10 minutes at 20°C
Recovery Time from Test: less than 5 seconds
Solution Level: 500ml
Calibration Stability: 1 year
FEATURES
DUPLEX Temperature Monitoring and Control System
Microprocessor controlled circuit to heat and maintain solution temperature of 34.00°C
Independent circuit for monitoring and display of solution temperature
Machined one piece Top Housing design with integrated manifold allows uniform heating of the inlet and outlet passages. This design allows for the elimination of condensation due to unheated or poorly heated tubing inside of the simulator
Controlled Head Space Heater (U.S. Patent No. 7,493,793). Rapidly heats top housing and passages to a uniform temperature eliminating equilibration time (15-60 minutes) usually needed once attaining 34.00°C.
Backlighted 132 x 32 pixel LCD Graphic Display
Temperature display resolution of .01°C
Visual indication of head space heater and solution heater activity.
Self diagnostics with audible and visual indication of system status. Status messages are displayed indicating exactly what is wrong with the simulator instead of error codes.
12 VDC operation
Included Power Supply – Input 100-240 V~1.5A, 47-63Hz; Output 12 VDC @ 8.33 A; UL/ cUL/ CE Listed
RS232 port for interfacing with popular breath analyzers
WARRANTY
One year, parts and laborstandard for testing or calibration of alcohol breath analyzers.